Water quality analysis and testing instrument of recirculating cooling water system in chemical palnt

The synopsis of project
The ERUN series water quality analysis and testing instrument of recirculating cooling water system in chemical palnt developed and produced by Erun Group is a solution designed in accordance with the GB50648. The chemical process cooling water quality testing instrument is suitable for testing the water quality of chemical industry recirculating cooling water using indirect open cooling systems, indirect closed cooling systems, direct open cooling systems in petrochemical, agricultural chemical, chemical pharmaceutical, polymer, coating, oil and fat, plastic, synthetic fiber, rubber, pharmaceuticals and other chemical industries. The main parameters that can be tested by chemical industry recirculating water quality analyzer include pH, conductivity, turbidity, suspended solids, total hardness, total alkalinity, chloride ions, iron, copper, ammonia nitrogen, COD, heterotrophic bacteria, oil, total phosphorus, residual chlorine, etc.
The contents of project

1、Classification of chemical industry:

The chemical industry can be divided into fuel chemical industry, inorganic chemical industry, organic chemical industry, and fine chemical industry.

(1)Fuel chemical industry

(2)Inorganic chemical industry

         a)Basic inorganic chemical industry

         b)Chemical mining industry

         c)Fertilizer and pesticide industry

(3)Organic chemical and fine chemical industry

         a)Organic synthesis industry

         b)Synthetic resin and plastic industry

         c)Rubber industry

         d)Chemical fiber industry

         e)Pharmaceutical and chemical industry

         f)Daily chemical industry

         g)Chemical additive industry

         h)Paint and dye industry

         i)Information materials industry

         j)Other chemical industries


2、Routine testing items of water quality analysis and testing instrument of recirculating cooling water system in chemical palnt:


Serial No.

Items

Indirect open cooling system

Indirect closed cooling system

Direct open cooling system

1

pH

Once a dayOnce a dayOnce a day

2

conductivity

Once a day

Once a day

Spot-check

3

turbidity

Once a day

Once a day

Once a day

4

suspended solids

Once to twice a month

Not tested

Once a day

5

total hardness

Once a day

Once a day or spot-check

Once a day

6

calcium hardness

Once a day

Once a day or spot-check

Once a day

7

total alkalinity

Once a day

Once a day or spot-check

Once a day

8

chlorine ions

Once a day

Once a day or spot-check

Once a day or spot-check

9

iron

Once a day

Once a day

Not tested

10

copper

Once a week

Once a week

Not tested

11

ammonia nitrogen

Once a weekOnce a weekNot tested

12

COD

Once a weekNot testedNot tested

13

heterophic bacteria

Once a weekOnce a weekNot tested

14

oil

Spot-checkNot testedOnce a day

15

drug concentration

Once a dayOnce a dayNot tested

16

total phosphorus

Once a dayNot testedNot tested

17

residual chlorine

Once a dayDepending on the medicationNo need for testing

18

biological clay content

Once a week

Once a week

Not tested

Note

1、The analysis of oil content is determined based on the specific situation of the indirect open cooling system in the refining unit.

2、The analysis of total phosphorus is applicable to phosphorus based formulation systems.

3、The analysis of Cu2+ and ammonia nitrogen is only used for copper heat exchanger systems.

4、The analysis of COD is applicable to systems that use recycled water as supplementary water or are prone to material leakage, and the analysis frequency can be appropriately increased based on changes in the water quality of the recycled water.

5、The testing method for the amount of biological sludge is the biological filter method, and the direct open cooling system does not need to be detected.


3、Technical Parameters of water quality analysis and testing instrument of recirculating cooling water system in chemical palnt:


Serial No.

Items

Measuring range

Unit

Maximum deviation

Analysis methods

Standard basis

1

pH

0.00-14.00

——

0.1±0.1

Glass electrode method

GB/T 6904

2

conductivity

0.00µS/cm-199.9mS/cm

mS/cm

±1.0% (F.S.)

Potassium chloride electrode method

GB/T 6908

3

turbidity

0-400

NTU

≤2%(F.S.)

Scattering method-formazine standard

GB/T 12151

4

suspended solids

5-1000

mg/L

≤2%(F.S.)

Light scattering method

——

5

total hardness

0.1-500.0

mg/L

5%(F.S.)

EDTA acid chrome blue K method

GB/T 6909

6

total alkalinity

0-1000

mg/L

5%(F.S.)

Potentiometric titration

GB/T 15451

7

chlorine ions

0.20-100.0

mg/L

5%(F.S.)

Ammonium thiocyanide titration method

GB/T 29340

8

iron

0.0-10.0

mg/L

5%(F.S.)

1,10-phenanthroline spectrophotometry

GB/T 14427

9

copper

0.02-10.0

mg/L

5%(F.S.)

Sodium diethyldithiocarbamate spectrophotometry

GB/T 13689

10

ammonia nitrogen

0.02-250.00

mg/L

3%(F.S.)

Nessler's reagent spectrophotometric method

HJ 535

11

COD

5-2000

mg/L

5%(F.S.)

Dichromate method

GB 11914

12

heterophic bacteria

——

个/L

——

Plate counting method

——

13

oil

0~900.00

mg/L

RSD<0.6%

Infrared photometry

GB/T 16488

14

total phosphorus

0.02-10.00

mg/L

5%(F.S.)

Ammonium molybdate spectrophotometric method

GB 11893

15

residual chlorine

0.02-10.00

mg/L

5%(F.S.)

Titrimetric method using N,N-diethyl-1,4-phenylenediamine

HJ 585


4、List of water quality analysis and testing instrument of recirculating cooling water system in chemical palnt:

Serial No.

Part Number

Name

Items

Pictures

1

ERUN-ST-MU97

Laboratory benchtop multi-parameter water quality analyzer

pH、conductivityturbiditysuspended solidstotal hardnesstotal alkalinitychlorine ionsironcopperammonia nitrogen、COD、total phosphorus、residual chlorine

化学工业循环冷却水化验室台式多参数水质测定仪


2

ERUN-ST-OD380

Laboratory benchtop infrared spectrophotometer

oil

化学工业循环冷却水化验室实验室台式红外分光测油仪


3

ERUN-SP-MD690

Portable water quality microbiological tester

heterophic bacteria

化学工业循环冷却水化验室便携式水质微生物检测仪

Note:Analysis parameter items can be flexibly combined and selected according to the customer's own needs.


5、Other parts list of water quality analysis and testing instrument of recirculating cooling water system in chemical palnt:

Serial No.

Name

Specifications

Quantity

1

Colorimetric tube

Ф25×95mm

10

2

Colorimetric tube

Ф16×110mm

10

3

Test tube rack

White

2

4

Pipette gun

Minimum graduation 0.1mL

2

5

Cuvette

50×25×70mm

2

6

Lens hood

Black

1

7

total hardness reagent

100 times/set

1

8

total alkalinity reagent100 times/set

1

9

chlorine ions reagent50 times/set

1

10

iron reagent50 times/set

1

11

copper reagent100 times/set

1

12

ammonia nitrogen reagent200 times/set

1

13

COD reagent

200 times/set

1

14

total phosphorus reagent100 times/set

1

15

residual chlorine reagent100 times/set

1

16

benchtop multifunctional CNC digestion instrument

ERUN-DIG-16B

1

17biochemical incubatorERUN-SHX1001
18temperature electrode/1
19pH composite electrode/1
20conductivity electrode/1
21chargers and charging cables/1
22USB cable/1
23instruction manual/1
24

Product certificate

/Product warranty card

/1

Note:Analysis parameter items can be flexibly combined and selected according to the customer's own needs.